Το έργο με τίτλο An external beam technique for proton-induced X-ray emission analysis από τον/τους δημιουργό/ούς Katsanos Anastasios, Xenoulis Alexander C., Hadjiantoniou A. , Fink Richard W. διατίθεται με την άδεια Creative Commons Αναφορά Δημιουργού 4.0 Διεθνές
Βιβλιογραφική Αναφορά
A. Katsanos, A. Xenoulis, A. Hadjiantoniou and R. W. Fink, "An external beam technique for proton-induced X-ray emission analysis", Nucl. Instrum. Meth., vol. 137, no. 1, pp. 119-124, Aug. 1976. doi: 10.1016/0029-554X(76)90256-1
https://doi.org/10.1016/0029-554X(76)90256-1
A method for proton irradiation of targets in free air is described. A 7.6 μm beryllium foil used as a proton exit window causes negligible energy loss for as low as 2 MeV protons and can withstand relatively high currents for extended periods of time without vacuum problems. Proton-induced X-rays from thin and thick samples are detected using Si(Li) and intrinsic Ge detectors A detailed comparison of X-ray spectra obtained both with external and internal bombardment, for a variety of samples, is presented and these results clearly demonstrate the superior performance of the external beam technique for elemental micro-analysis.