Design of a Microprocessor-Based Sphygmomanometer.

Kalaitzakis Kostas, Papamarkos, N, Vachtsevanos, G.J

Full record


URI: http://purl.tuc.gr/dl/dias/0ED259C0-EFC9-4BB5-B6A4-61BEDD8171EF
Year 1990
Type of Item Peer-Reviewed Journal Publication
License
Details
Bibliographic Citation K.C. Kalaitzakis, N.E. Papamarkos, and G.J. Vachtsevanos, "Design of a Microprocessor-Based Sphygmomanometer," Biomedical Instrumentation and Technology, vol. 24, pp. 31-36, Jan-Feb. 1990
Appears in Collections

Summary

This paper describes the implementation on a microprocessor of a new method for the indirect measurement and recording of the systolic and diastolic blood pressure in humans. The technique is based on a statistical analysis of the cardiac pulse pressure signal. Polynomial relations are derived between the amplitude of the pulsatile pressure waveforms at the systolic and diastolic points and the amplitude of pulse signals detected when the artery is fully occluded. With the dual objective of automating the measurement procedure and minimizing errors, an electronic analog-digital sphygmomanometer that contains suitable electronic instrumentation was developed. The functions of processing the pressure signal, automating the measurement, and recording the results are performed and controlled by a microprocessor. A laboratory prototype embodying this approach was constructed and its performance and reliability were verified using a series of clinical tests. The test results indicate that the device is accurate within acceptable bounds for automated blood pressure instruments.